FACTORY MENU > UTILITIES > SIGNAL ROUTING
Note that access to the UTILITIES requires DIP2 = ON at the mainboard.
TRANSM. LOAD mA shows the load of the transmitter coil to the mainboard transmitter driving circuit.
Typical load mA is 100 to 150 mA (all frequencies, all sensor types)
In case the load is much higher, changing the frequency Hz could sometimes help to reduce the load. Change the frequency Hz in 1000 Hz steps up and down, see if this reduces the mA load.
Frequency is found here:
FACTORY MENU > SENSOR CALIB. > FREQUENCY x[Hz]